Laboratory Geometric Calibration of Non-metric Digital Camera

YUAN Feng1, ZHANG Jixian2, QI Weijun1,2, FANG Aiping1,2   

  1. 1. National Calibration Center for Electro-optical Distance Meter, Beijing 100039, China;
    2. Chinese Academy of Surveying and Mapping, Beijing 100830, ChinaAbstract
  • Received:2014-11-13 Revised:2016-02-03 Online:2016-05-20 Published:2016-05-30
  • Supported by:
    Public Science Research Program of Surveying, Mapping and Geoinformation (No. 201512010);Chinese Academy of Surveying and Mapping Special Fund (No. 7771531)

Abstract: Geometric calibration of non-metric digital camera, including the accurate determination of interior orientation elements and distortion parameters, directly affects the ultimate accuracy of its using in aerial photogrammetry. In this paper, a laboratory geometric calibration system of non-metric digital camera is developed. A collimator and a star tester are used as the target generator in this system. After measuring locations of targets in CCD and corresponding angles of parallel lights, high precision geometric calibration can be accomplished. Two kinds of mainstream digital aerial cameras, Canon EOS 5D Mark Ⅱ and Hasselblad H3D, are used to take experiments based on this system and the results are analyzed. Experiment results show that the method of this paper is correct and the calibration precision is reliable.

Key words: non-metric digital camera, geometric calibration, interior orientation elements, distortion

CLC Number: